Author:
Publisher:
ISBN:
Category :
Languages : en
Pages : 15
Book Description
Local measurements of n{sub e} and T{sub e} in the divertor region are necessary for a more complete understanding of divertor physics. We have designed an extension to the existing multipulse Thomson scattering system to measure n{sub e} in the range 5 x 1018 to 5 x 102° m−3 and T{sub e} 5--500 eV, with 1 cm resolution from 1--21 cm above the floor of the DIII-D vessel, in the region of the X-point for lower single-null diverted plasmas. One of the existing 8, 20 Hz, ND:YAG lasers will be redirected to a separate vertical port, and viewed radially with a specially designed, f/6.8 lens. Fiber optics carry the light to additional polychromators whose interference filters have been optimized for low T{sub e} measurements. Other aspect of the system, including the beam path to the vessel, polychromator design, real time data acquisition, laser control, calibration facility, and DIII-D timing and data acquisition interface will be shared with the existing multipulse Thomson system. An in-situ laser alignment monitor will provide alignment information for each laser pulse.